Elektrodengläser mit Zusatz von Germaniumdioxyd

Teil III

Authors

  • Dorothée Wegmann Organisch-Chemisches Laboratorium der Eidgenössischen Technischen Hochschule, Zürich
  • W. Simon Organisch-Chemisches Laboratorium der Eidgenössischen Technischen Hochschule, Zürich

DOI:

https://doi.org/10.2533/chimia.1962.161

Abstract

The alkaline error at various pH values (0.15 and 1.0N Na+) and the water solubility of some electrode glasses have been determined. The behaviour of these electrode glasses after prolonged treatment with 1.0 N NaOH at 75 °C has been studied. It is found that the substitution of 1 mole-% silicon dioxide by germanium dioxide, which improves the workability of electrode glasses, does not significantly influence their alkaline error, water solubility and behaviour under extreme conditions.

Downloads

Published

1962-05-31

Issue

Section

Kurze Mitteilungen

How to Cite