Einsatz einer Mikrowaage zur Messung der Gassorption an Katalysatoren
DOI:
https://doi.org/10.2533/chimia.1980.181Abstract
The use of microbalances to study gas-solid reactions at high temperatures and normal pressure is often limited by fluctuations due to thermal convection in the reaction zone. The observed fluctuations depend on the pressure, the type of gas, the temperature and the geometry of the arrangement. The problems have been overcome by distributing the gas on the circumference of the tube surrounding the specimen. The observed fluctuations under normal pressure (1,1 bar) and temperatures up to 200 °C were less than 0,5 μg.
Downloads
Published
1980-04-30
Issue
Section
Kurze Mitteilungen
License
Copyright (c) 1980 Beat Walker

This work is licensed under a Creative Commons Attribution 4.0 International License.
How to Cite
[1]
B. Walker, M. Dettmer, A. Renken, Chimia 1980, 34, 181, DOI: 10.2533/chimia.1980.181.